跳转到主要内容

The Picarro SI2103 gas concentration analyzer delivers ultraprecise, parts-per-trillion (ppt) measurements of ammonia (NH3) in a reliable, easy-to-use, compact design. The SI2013 can be commissioned and operating within minutes, requires no field calibration or consumables, and can operate for months without user interaction.

  • Real-time AMC monitoring in cleanrooms, FOUP and fab equipment
  • Fast, continuous analysis in seconds
  • Virtually no downtime or consumables cost
  • Free of interference from common fab chemicals
  • Extreme accuracy for event confirmation
  • No field calibration required

Gas Detected

Ammonia (NH3)

Precision (1σ)

≤300 ppt (10 sec),
≤100 ppt (100 sec)

Lower Detectable Limit (3σ)

300 ppt

Method Detection Limit
(per Semi C10-1109)

500 ppt

Linearity
(per IEC 61207)

±1%

Accuracy at span

±5% @ full scale

Accuracy at zero

±100 ppt

Instrument-to-Instrument Consistency

±5% @ full scale
±100 ppt @ zero

Measurement Range

0-10 ppm

Measurement Interval*

<4 seconds

Sample Flow Rate

~2 slm

Combined Response Times
(T90/10 + T10/90) @ 20 ppb

<3 minutes

Fall Times
T90/10 @ 20 ppb

<1 minute

*-Measurement interval at span may increase as much as 2x above listed values.

Measurement Technique

Cavity Ring-Down Spectroscopy

Calibration Period Recommendation

Calibration not required - Initial validation at 6 months, then every 12 months thereafter

Time Required to Perform Validation

Estimated at <15 minutes per manufacturer's instructions

Measurement Cell Temperature Control

±0.005°C

Measurement Cell Pressure Control

±0.0002 atm

Operating Temperature

15 to 35°C (operating); -10 to 50°C (storage)

Ambient Humidity

<99% RH non-condensing

Accessories

Pump (external, included), keyboard (included), mouse (included), LCD monitor (optional), Maintenance Kit (optional)

Communication Interfaces

RS-232, Ethernet, USB, analog 0–10 V, Modbus, 4-20mA (optional)

Sample Inlet Connection

1/4" Stainless Steel Swagelok® Tube Fitting (recommended 1/4" OD PFA Tubing)

Dimensions

Analyzer: 17" w × 8.38" h x 24.4" d (43.2 × 21.3 × 62 cm)

Weight

73 lbs. (33.18 kg) including external pump

Power Requirements

100–240 VAC, 47–63 Hz (auto-sensing), <400 W (total): 250 W (analyzer), 150 W (pump) at steady state

Warranty

12 months

Certifications

CE conformance

Country of Manufacture

USA

Components/Parts

Actions

Frequency

Particulate filter

Replace

Every 6 months

Measurement Validation 

Proxy validation 

Every 12 months

Pump

Rebuild (diaphragm)

Preventive maintenance every 2 years

Fan assemblies  

Replace

Preventive maintenance every 2 years

External Pump 

Replace

Every 5 years