The SAM-S AMC Mobile Monitoring System integrates Picarro’s industry-leading cavity ring-down spectroscopy (CRDS) analyzers into a high-performance sampling system. Traditional sampler designs degrade individual analyzer performance through the use of a linear manifold which limits the gas flow rate to each analyzer. As a result, traditional samplers relegate analyzer performance to the lowest common denominator. Time to detect an AMC is ultimately affected, which can lead to a false alarm or even missing an AMC event altogether.

The superior design of the SAM system is optimized to ensure the highest combined performance, utilizing a patent-pending non-linear multiplexing system that enables high gas flow rates, minimizes cross port contamination, and quickly reports accurate contaminant concentrations. The Picarro SAM-S system can be configured to sample multiple locations, accommodate up to two of Picarro’s SI series of analyzers to measure a variety of contaminants. 

Gases Detected

NH3, HF, HCl, SO2, H2S (depending on choice of up to 2 analyzers from table under Configurations tab)

Sampling Line

1/2"OD x 3/8" ID UHP-PFA tubing

Dimensions

System: 33" (W) x 33" (D) x 48" (H) (84 x 84 x 122 cm)
Monitor Height: 12.5" (32 cm)

Weight

8 Port: 405 lbs (183 kg)

16 Port: 430 Ibs (195 kg)

Power Requirements

220-240 VAC single phase, 50/60 Hz, 15 Amp

Communication

Ethernet TCP/IP, RESTful API

Warranty

12 months

Certifications

CE, SEMI S2

Country of Manufacture

USA

Analyzers

Gases Measured

SI3401

NH3, HF, HCl

SI2205

HF

SI2108

HCl

SI2104

H2S

SI2306

HF, NH3

SI2103

NH3

SI5450

SO2

Part Numbers (Other Combinations Available)

SAM-S-08-U-AO

8 port AMC system for NH3, HF, and HCl with UPS (Uninterrupted Power Supply)

SAM-S-16-N-A0

16 port stationary AMC system for NH3, HF, and HCl

SAM-S-16-N-AG

16 port stationary AMC system for NH3, HF, HCl, and SO2

Components/Parts

Actions

Frequency

Zero filter

Replace

Every 6 months

Pump

Replace

Every 1 year

Fan assemblies  

Replace

Every 2 years